Freon Improves Laser-Ablation Spectrometry
To reduce the incidence of elemental fractionation and improve the accuracy of laser ablation induction-coupled plasma mass spectrometry for the isotopic analysis of geological samples, Takafumi Hirata of
Tokyo Institute of Technology suggests the addition of a refrigerant. In the Jan. 15 issue of
Analytical Chemistry, the researcher reported that the delivery of Freon gas to laser ablation sites improves the precision of measurements of the
206Pb/
238U ratio in ancient rocks.
Typically, the laser systems increase the Pb/U ratio by 20 to 50 percent as measurement progresses. Hirata found that exposing the ablation sites to 60 µl/min of Freon gas during the process offered a precision of 2 to 4 percent for the 1 billion-year-old and 2.4 billion-year-old zircon standards used in the experiment.