Search
Menu
Vescent Photonics LLC - Lasers, Combs, Controls 4/15-5/15 LB

Thermal Issues Associated With Flow Cytometry Miniaturization Solved

Facebook X LinkedIn Email
Positioning equipment supplier Siskiyou Corp. said it recently helped another manufacturer miniaturize its flow cytometer by developing adjustable mounting hardware that conducts heat away from the system's light source.

The manufacturer had the goal of reducing the length, complexity and cost of beam setup to improve instrument stability and reliability. The desired miniaturization also aimed to allow for the system's laser excitation modules to be placed in closer proximity to the flow cell.

Flow cytometers have inherent trade-offs between speed and sensitivity. Higher sensitivity is achieved by restricting the diameter of the core stream to just a few microns through a process called hydrodynamic focusing. The excitation laser beam is then focused down to a similar diameter, resulting in a very bright signal from each cell. However, a narrower core stream reduces the rate at which the sample passes through the flow cell.

This slower speed can be particularly problematic when counting very rare blood cells. This makes it critical that this higher-sensitivity configuration involves no further performance trade-offs or compromises. Furthermore, since micron shifts in a focused beam position affect overall detection efficiency, proper laser alignment is more critical in a narrow core system.

This typically makes these high-sensitivity setups less robust than conventional instruments with wider streams, thus necessitating a higher level of maintenance. In fact, optical misalignment is a leading cause of performance degradation in these instruments and is a common reason for field service. By physically placing the laser or lasers closer to the interaction zone, the optical throw distance is reduced, thus creating a more stable and reliable final product.

The cytometer manufacturer contracted Diode Laser Concepts Inc. (DLC) of Medford, Ore., a supplier of custom OEM laser-diode-based solutions, to design a new red (640-nm) laser engine for its miniaturized cytometer. DLC then turned to Siskiyou to create the adjustable mounting hardware.

BAE Systems Sensor Solutions - Fairchild - FS Sensor Solutions 4/24 MR

This application required a mounting arrangement that would allow for 3D, linear (XYZ) adjustment of the focused laser spot. Key design choices were selecting a stage material and bearing type that would achieve the necessary adjustment range and resolution, as well as to provide lockable, long-term mechanical stability while meeting the overall package size and cost constraints.

In this case, 20-mm travel cross roller bearing aluminum stages with 100-threads-per-inch adjustability were chosen, since these standard products deliver the requisite performance in the most cost-effective format.

The laser diode is thermoelectrically cooled to maintain constant output characteristics, such as beam pointing, beam shape and wavelength. The heat must be efficiently coupled through the metal stages to the base platform, which acts as the final heat sink. While better heat flow enables the instrument to more rapidly achieve thermomechanical stability at the submicron level, this thermal management should not unduly impact cost or complexity.

After several iterations, Siskiyou found a solution based on connecting three stages via customized thin aluminum plates. Testing at DLC showed that this approach achieved submicron-level repeatability and stability at the flow cell, with a warm-up time of less than 1 minute. For perspective, it is not uncommon for large-frame flow cytometers to require a 30-minute warm-up time for the laser and beam delivery setup, which has a significant, negative impact on measurement costs and laser lifetime.

Approved by the flow cytometer manufacturer, the subsystem is now incorporated in the company’s next-generation instrument.

For more information, contact Mike Robinson at [email protected] and John Wingerd at [email protected].


Published: August 2015
Glossary
flow cytometry
Flow cytometry is a powerful technique used in biology and medicine for the quantitative analysis of the physical and chemical characteristics of cells and particles suspended in a fluid. The method allows for the rapid measurement of multiple parameters simultaneously on a cell-by-cell basis. It is widely used in various fields, including immunology, microbiology, hematology, and cancer research. Here are the key components and features of flow cytometry: Sample preparation: Cells or...
positioning
Positioning generally refers to the determination or identification of the location or placement of an object, person, or entity in a specific space or relative to a reference point. The term is used in various contexts, and the methods for positioning can vary depending on the application. Key aspects of positioning include: Spatial coordinates: Positioning often involves expressing the location of an object in terms of spatial coordinates. These coordinates may include dimensions such as...
Businessdiode lasersAmericasSiskiyoudiode laser conceptsDLCflow cytometryLaserspositioningBiophotonicsImagingBioScan

We use cookies to improve user experience and analyze our website traffic as stated in our Privacy Policy. By using this website, you agree to the use of cookies unless you have disabled them.